Reference Help and Troubleshooting & Support 17
Troubleshooting and Support
Agilent Aria Real-Time PCR Program 368
No amplification plots are visible in
the Graphical Displays screen
This could occur if the wrong data collection point was assigned to the
amplification plots. Reset the data collection point for the amplification
plots on the Analysis Criteria screen.
A lack of amplification plots would also occur if no data collection point
was set in the amplification segment on the Thermal Profile Setup screen.
Without a data collection point, no data is collected and analysis is
impossible.
You will also not see any amplification plots if you selected ΔRn data when
a reference dye was not defined for the experiment.
Signal fluctuations in amplification
plots
Possible sources of signal noise include environmental factors such as
vibration of the bench from other instruments, direct sunlight falling on the
back of the instrument, or fluctuations in the line power. These problems
can normally be resolved by relocating the instrument to a different bench
in the lab or, for the case of line power problems, by connecting the
instrument to a Power Conditioner or UPS.
In amplification plots, the baseline
fluorescence signals gradually
decrease across the baseline cycle
range
To reduce this effect, enable the instrument’s Warm Up feature for the
optical modules, which was implemented in version 1.8 of the instrument
firmware. From the touchscreen Home screen, go to Settings > System
Settings > Warm Up Settings. On the Warm Up Settings screen, select
Yes and press OK. The system turns on the optical module LED lights so
that the modules are fully warmed up prior to starting an experiment. The
warm up is repeated each time the instrument is powered on (after
completion of health checks) and after the optical modules door is closed.
The process takes approximately 15 minutes.
This feature is not recommended as a standard practice for all assays. In
some circumstances, however, warming up the optical modules may help
signals remain steady throughout the baseline cycle range. Consult with
Technical Support if you have questions about this feature.
Low increase in fluorescence with
cycling
The probe is not binding the target efficiently. Lower the annealing
temperature and verify the melting temperature.
Target PCR product is too long; redesign primers to yield a PCR product
<150 bp in length.
Magnesium concentration is too low; run a titration to optimize
concentration.
Insufficient or non-specific product is being formed. Verify product
formation through gel electrophoresis.